Mavolux 5032 C
الوصف
Product information:
With accuracy according to class C, the industrial luxmeter Mavolux 5032 C USB is primarily used for general applications. The smallest of four measuring ranges begins with an initial sensitivity of 0.1 lx. V(λ) adaptation deviation, which amounts to f1‘< 7.5 %, is considerably better than the admissible error limit for class C.
With the help of an optional luminance attachment with an acceptance angle of 15°, the meter can be used as an unclassified luminance meter. Luminance is measured in candelas per square meter (cd/m²) or footlamberts (fL), and defines the perceived brightness of a back-lighted or reflecting surface.
The manufacturer will provide a factory calibration certificate or a DAkkS calibration certificate which can be ordered separately. Depending on how the meter is used, the manufacturer recommends a calibration interval of 12 to 24 months.
Measuring Functions of Mavolux 5032 C:
| Illuminance (LUX): | 0.1lx… 199,900 lx/
0.01 fc …19,900fc |
| Luminance: | 1cd/m²… 1 999 000 cd/m²/0.1fL with optional luminance attachment, not classified |
| Measuring Ranges: | 4 |
| Measuring Range Selection: | Automatic / manual |
| Measuring Rate: | 2 per second |
| Measuring Method: | Distance measurement, Contact measurement with optional luminance attachment |
| Measuring Sensor: | Silicone photodiode with V (λ) filter |
| Measurement Cable: | 1.5m firmly connected |
| Measured Value Memory: | 100 measured values |
| Error Limit V(λ) adapted (f1’), typical: | <7.5% |
| Error Limit – overall error, typical: | ≤15% |
| Accuracy: | ± 3% of reading ± 1 digit |
Operation:
| Display: | 3 ½ digit LCD |
| Operating Elements: | 6 Keys |
| Interface: | USB 1.1 |
| Software: | GLUX 2 |
Miscellaneous:
| Operating Temperature: | 0°C to 50°C |
| Dimensions: | 65mm×120mm×19mm (meter)
31mm×105mm×30mm (probe) |
| Weight: | 200g (without battery) |
Delivered complete with Aluminium case, GLUX 2 software, USB cable, battery, operating instructions, calibration protocol